Product Description
This equipment contains button batteries and is not intended for general consumer use.
(i) the equipment is intended to be used in trades, professions or industries only;
(ii) the equipment is not intended for sale to the general public;
(iii) the equipment is not intended to be used where children are present;
Formerly Horiba B771
Determine water conductivity with as little as 0.12 mL of sample.
The conductivity of rain water is a trusted guide to determining atmospheric purity. In agriculture, measuring the conductivity of soil allows farmers and agronomists to determine optimum fertiliser usage and check the ‘health’ of soil after salt water damage. The LAQUAtwin meter makes conductivity testing simple, anywhere.
Applications include:
Fresh water testing (rain, rivers, lakes, hot springs); aquaria; soil testing for improved agriculture and analysis of salt water damage; analysis of surface cleanliness prior to coating; and improved paint adhesion.
Functions
- Salt/TDS Measurement
- Auto range change
- Temperature conversion(2%/℃ fixed)
- IP67 waterproof
- Auto hold
- Automatic power off(15 minutes)
Also known as:
Conductivity meter, EC meter, compact conductivity meter, compact EC meter, pocket conductivity meter, pocket EC meter, HORIBA EC meter, LaquaTwin conductivity meter, waterproof conductivity meter
MODEL
|
EC 11 |
EC 22 |
EC 33
|
Measurement principle
|
2 AC bipolar |
Minimum
sample volume
|
0.12mL or more
|
Measurement range
|
Conductivity:0 to 19.9mS/cm |
Conductivity:0 to 199mS/cm |
Conductivity:0 to 199mS/cm
TDS:0 to 9900ppm
|
Range and Resolution
(Valid numbers)
|
① 0 to 1999 μS/cm : 1μS/cm
② 2.0 to 19.9mS/cm : 0.1mS/cm
|
① 0 to 1999 μS/cm : 1μS/cm
② 2.0 to 19.9mS/cm : 0.1mS/cm
③20 to 199mS/cm : 1mS/cm*1
|
① 0 to 1999 μS/cm : 1μS/cm
② 2.0 to 19.9mS/cm : 0.1mS/cm
③20 to 199mS/cm : 1mS/cm*1
|
Calibration
|
Two-point*2 |
Three-point*2 |
Three-point*2
|
Accuracy
|
±2%F.S.±1digit (0 to 199 μS/cm ) |
±2%F.S.±1digit (0 to 199 μS/cm )
±5%F.S.±1digit (20 to 199mS/cm) |
±2%F.S.±1digit (0 to 19.99 μS/cm )
±5%F.S.±1digit (20 to 199mS/cm)
|
Functions
|
Salt/TDS Measurement・Auto range change・Temperature conversion(2%/℃ fixed)・Waterproof*3・Auto hold・Automatic power off(15 minutes)
|
Auto range change・Temperature conversion(2%/℃ fixed)・Waterproof*3・Auto hold・Automatic power off(15 minutes) |
TDS Measurement・Auto range change・Temperature conversion(2%/℃ fixed)・Waterproof*3・Auto hold・Automatic power off(15 minutes) |
Display
|
Custom (monochrome) Digital LCD
|
Operating
temperature/humidity
|
0 to 50˚C, 85% or less in relative humidity (no condensation) |
Power
|
CR2032 batteries (x2) |
Battery life
|
Approx. 400 hours in continuous use |
Main Material
|
ABS epoxy |
Dimensions
/Mass
|
164mm × 29mm × 20mm (excluding projections) / Approx. 45 g (meter only, without batteries) |
Accessories included
|
2 CR2032 batteries/1 Pipette/Instruction manual/Quick manual/Storage case/Standard solution(1.41mS/cm) (14mL)/Treatment reagent (14mL)
※For the high conductivity standard solution (12.9mS/cm) is sold separately. |
*1 When the measured value is out of the measurement range, the displayed value blinks. It should be used only as a guide.
*2 Selectable between one-point and two-point calibrations. High conductivity standard solution (12.9mS/cm) is sold separately.
*3 IP67:no failure when immersed in water at a depth of 1 meter for 30 minutes. But the product can not be used underwater.