Portable Meter Kit (PH/ORP/DO/Temp) LAQUA PD210-K

From: $1,400.00 ex GST

Dual Channel Meter and Includes 9652-10D Plastic-Body Gel-filled pH/Temp Electrode with 2m Cable, 9552-20D Galvanic DO/Temp probe with 2m Cable, pH 4, 7, 10 Buffers (60ml each), Electrode Adapter and Carrying Case
Usually Stocked
Note: ORP electrode not included. ORP electrode 9300-10D to be purchased separately.

ORP Electrode 9300-10D

Usually stocked

Product Description

Features

Note: ORP electrode not included. ORP electrode 9300-10D to be purchased separately.

 

  • Automatic pH calibration – up to 5 points (NIST, USA pH buffers) / up to 6 points (DIN pH buffers)
  • 0.01 pH resolution
  • Offset and average slope display
  • Automatic / manual temperature compensation and temperature calibration function
  • ORP calibration option
  • Dissolved oxygen (DO) concentration in ppm (mg/L) and % saturation
  • 0.01 mg/L or 0.1% DO resolution
  • Compensates salinity, temperature, and atmospheric pressure
  • Up to 2 DO calibration points
  • Auto log data; built-in memory can store up to 500 data
  • Auto shut-off (programmable: up to 30 minutes)
  • Auto stable, auto hold, and real time measurement modes
  • Calibration alarm (programmable: up to 90 days)
  • Built-in electrode holder allows 2 electrodes to be attached
  • Monochrome LCD with white LED backlight: 50 x 50 mm
  • Phono jack output for connecting the meter to computer for software upgrade
  • Casing is IP67 waterproof/dustproof, shock & scratch resistant, non-slip
  • Powered by 2 x AA batteries
  • Packaged with electrodes and calibration solutions in carrying case
  • 3 years meter warranty

Meter Kit

PD210-K

Package Details:

  • LAQUA PD210 dual channel pH/ORP/DO/Temp meter
  • 2 x AA batteries
  • Manual and quick guide
  • 9652-20D plastic-body, gel-filled pH electrode with integrated temperature sensor, 2m cable
  • 9552-20D Galvanic DO probe with integrated temperature sensor, 2m cable
  • pH 4.01, 7.00, 10.01 buffers (60ml each)
  • Electrode adapter
  • Carrying case

Note: ORP electrode not included. ORP electrode 9300-10D to be purchased separately.

PD210-KN: Kit with pH 4.01 / 6.86 / 9.18 NIST buffers (60ml each) is available upon request.

Manufactured by HORIBA

Additional Information

Weight 1 kg
Dimensions 1 × 1 × 1 cm

Model

LAQUA PD210

pH/ORP/DO/Temperature (°C/°F)

pH Range

-2.00 to 16.00 pH

Resolution

0.01 pH

Accuracy

± 0.01 pH

Calibration Points

Up to 5 (USA & NIST) / Up to 6 (DIN) 

Buffer Options

USA, NIST, DIN

ORP Range

± 2000 mV

Resolution

0.1 mV (< ±1000 mV), 1 mV (≥ ±1000 mV)

Accuracy

±0.3 mV (< ±1000 mV), 0.3% of reading (≥ ±1000 mV)

Calibration Option

Yes

DO Range

0.00 to 20.00 mg/L / 0.0 to 200.0 %

Resolution

0.01 mg/L or 0.1 %

Accuracy

±0.1 mg/L

Salinity Compensation

0.0 to 40.0 ppt

Barometric Pressure Compensation

Yes

Temperature Compensation

Yes

Calibration Points

Up to 2

DO Probe Type 

Galvanic integrated with temperature sensor

Temperature Range

-30.0 to 130.0 °C / -22.0 to 266.0 °F

Resolution

0.1 °C / °F

Accuracy

± 0.5 °C / ± 0.9 °F

Calibration Option

Yes

Memory Data

500

Auto Log Data

Yes

Measurement Modes

Auto Stable, Auto Hold, Real Time

Offset & Average Slope Display

Yes 

Calibration Alarm

Yes (Programmable: up to 90 days)

Auto Shut-Off

Yes (Programmable: up to 30 minutes)

Electrode Status

Yes (On screen display)

Diagnostic Messages

Yes

Software Upgrade

Yes via PC (USB) cable PN 3200779639

Meter Inputs

2 x BNC (for pH and DO electrodes), 2 x phono (for temperature sensors)

Meter Output

phono (for software upgrade only)

Display

Custom LCD with backlight 

Housing

IP67, shock & scratch resistant, non-slip

Power Requirement

2 x AA batteries

Battery Life

> 500 hours

Meter Dimensions

160 (L) x 80 (W) x 40.60 (H) mm

Meter Weight

Approx. 260g (with batteries) / 216g (without batteries)

Click Here for LAQUA 200 Series Standard Handheld Water Quality Meters Brochure

Click Here for LAQUA PD210/220 Instruction Manual

Click Here for pH Meter Quick Guide

Click Here for Dissolved Oxygen Meter Quick Guide

CLICK HERE FOR AUTO STABLE MEASUREMENT MODE

CLICK HERE FOR AUTO HOLD MEASUREMENT MODE

CLICK HERE FOR REAL TIME MEASUREMENT MODE

CLICK HERE FOR pH CALIBRATION

CLICK HERE FOR VIEWING pH CALIBRATION DATA

CLICK HERE FOR ORP CALIBRATION

CLICK HERE FOR TEMPERATURE CALIBRATION

CLICK HERE FOR SETTING CELL CONSTANT

CLICK HERE FOR CONDUCTIVITY CALIBRATION

CLICK HERE FOR VIEWING CONDUCTIVITY CALIBRATION DATA

CLICK HERE FOR DISSOLVED OXYGEN (DO) CALIBRATION

CLICK HERE FOR VIEWING DISSOLVED OXYGEN (DO) CALIBRATION DATA

keyboard_arrow_up